The CSIM11 measures pH for the range of 0 to 14 pH with an accuracy of ±0.1 pH unit. The sensor is designed for submersion or insertion into tanks, pipelines, and open channels. The sensor features a plunger-style pH glass electrode allowing it to be mounted at any angle. The porous Teflon® liquid junction is less susceptible to clogging as compared to conventional reference junctions. A titanium ground rod runs inside the PPS outer body to eliminate ground loop errors. An internal amplifier boosts the signal, decreasing signal interference.
Features
0 to 14 pH measurement range; accuracy is ±0.1 pH unit
Operable environments are 0° to +70°C; pressure 0 to 30 psig
Requires 107 probe for temperature measurement and compensation
Manufactured by Innovative Sensors
Specifications
pH Range: 0 to 14
Zero Potential: 7.0 pH ±0.2 pH
Sodium Error: <0.05 pH in 0.1 Molar Na+ ion at 12.8 pH
Output: ±59 mV/pH unit
Pressure Range: 0 to 30 psig
Accuracy: ±0.1% over full range
Impedance: <1 Mohm @ 25°C
Reference Cell: Single Junction KCl/AgCl
Response Time: 95% of reading in 10 s
Drift: <2 mV per week
Producing Area:USA