The D701 is a polarographic O2 electrode (identical in design to the E-5046 electrode from Radiometer) consisting of a combined platinum- and silver/silver chloride electrode mounted in an electrode jacket and covered by a 20 um thick membrane of polypropylene. A reservoir of phosphate buffer and potassium chloride is trapped inside the electrode.
When a polarizing voltage of approximately 630 mV is applied to the platinum cathode, a current proportional to PO2 will be generated by reduction of the oxygen diffusing towards the cathode.
SPECIFICATIONS
Range 0 - 800 mm Hg
Response Time 30 - 60 sec. at 37oC (99% response)
Oxygen Sensitivity 10-11 A / mm Hg PO2 (approx.)
Zero Current Less than 3 x 10-10 A at 37oC
Accuracy Approx. 1%
Temperature Range 20 - 50oC
Temperature Coefficient Approx. 3% / oC when equipped with a 20 um polypropylene membrane
Reference Electrode Silver/silver chloride
Membrane Material polypropylene, thickness 20 um
Electrode Solution Phosphate buffer - potassium chloride
Plug type BNC UG-88/u
Cable Length 1 m
Electrolyte Reservoir Approx 0.8 mL
INCLUDED SUPPLIES:
Electrode body & O-ring
Oxygen membranes (15/pkg)
Membrane O-ring (10/pkg)
Membrane mounting tool
Oxygen internal fill solution (electrolyte) (60 mL)
Rubber end cap
Producing Area:Denmark